Specification-31 
4 
Service 
TABLE 
1-1 
(cont) 
Capacitively Coupled (AC) 
Supplemental Information 
- 
- 
- 
Characteristic 
Maximum Input Voltage 
Direct Coupled (DC) 
lnput R and C 
Resistance 
-- 
-- 
- 
- 
- 
- 
Performance Requirement 
Without Probe 
I 
With P6149  Probe 
1 Mn, within 2%  10 Mn, within 2% 
300 V, dc plus 
peak ac 
Capacitance 
1 
47 pF. within 4 pF  13.5 pF or less 
1 
500 V, dc plus 
peak ac 
Signal Delay  Permits viewing leading edge of 
waveform that initiates trigger 
HORIZONTAL SYSTEM 
Chopped Repetition Rate 
Sweep 
Time/Div 
Calibrated Range 
Accuracy 
Unmagnified 
5 s to 0.5 
s/Div 
0.2 s to 1 ps/Div 
Magnified 
0.5 s to 50 ms/Div 
20 ms to 0.5 ps/Div 
0.2 ps and 
0.1 pslDiv 
TIME/DIV Variable 
1-2 
Approximately 100 kHz 
5 s to 1 ps/Div. XI0 magnifier 
extends sweep time/div to 0.1 ps 
Within 4%  over center 
8 
graticule 
divisions. Within 5% over any 
2 
division interval within center 
8 
divisions. 
Within 3% over  center 
8 
graticule 
divisions. Within 4% over any 2 
division interval within center 
8 
divisions. 
Within 5% over center 
8 
graticule 
divisions. Within 6% over any 2 
division interval within center 
8 
divisions. 
Within 4%  over center 
8 
graticule 
divisions. Within 5% over any 
2 
division interval within center 
8 
divisions. 
Within 5% over center 
8 
graticule 
divisions. Within 6% over any 2 
division interval within center 
8 
divisions, excluding first and 
last 2 divisions of magnified sweep. 
At  least 2.51 
21 steps in a 1-2-5 sequence 
REV. 
A, 
AUG. 
1977