Specification-31
4
Service
TABLE
1-1
(cont)
Capacitively Coupled (AC)
Supplemental Information
-
-
-
Characteristic
Maximum Input Voltage
Direct Coupled (DC)
lnput R and C
Resistance
--
--
-
-
-
-
Performance Requirement
Without Probe
I
With P6149 Probe
1 Mn, within 2% 10 Mn, within 2%
300 V, dc plus
peak ac
Capacitance
1
47 pF. within 4 pF 13.5 pF or less
1
500 V, dc plus
peak ac
Signal Delay Permits viewing leading edge of
waveform that initiates trigger
HORIZONTAL SYSTEM
Chopped Repetition Rate
Sweep
Time/Div
Calibrated Range
Accuracy
Unmagnified
5 s to 0.5
s/Div
0.2 s to 1 ps/Div
Magnified
0.5 s to 50 ms/Div
20 ms to 0.5 ps/Div
0.2 ps and
0.1 pslDiv
TIME/DIV Variable
1-2
Approximately 100 kHz
5 s to 1 ps/Div. XI0 magnifier
extends sweep time/div to 0.1 ps
Within 4% over center
8
graticule
divisions. Within 5% over any
2
division interval within center
8
divisions.
Within 3% over center
8
graticule
divisions. Within 4% over any 2
division interval within center
8
divisions.
Within 5% over center
8
graticule
divisions. Within 6% over any 2
division interval within center
8
divisions.
Within 4% over center
8
graticule
divisions. Within 5% over any
2
division interval within center
8
divisions.
Within 5% over center
8
graticule
divisions. Within 6% over any 2
division interval within center
8
divisions, excluding first and
last 2 divisions of magnified sweep.
At least 2.51
21 steps in a 1-2-5 sequence
REV.
A,
AUG.
1977