Section 
1-314 
Service 
SPECIFICATION 
The  Sony-Tektronix  314  is  a  dual-channel,  10 
megahertz, portable storage oscilloscope. The instrument 
is constructed to withstand a wide range of environmental 
conditions. 
Dc  to  10  megahertz  vertical  bandwidth  provides 
calibrated  deflection  factors  from  1  millivolt  to  10 
volts/division. 
The horizontal system provides calibrated sweep rates 
from  5  seconds  to  1  microsecond/division  and  a XI0 
sweep magnifier to increase the fastest sweep rate to 0.1 
microsecond/division. Three sweep modes are available: 
AUTO,  NORM, and  SINGLE  SWP.  The trigger  circuit 
provides a stable display over the full bandwidth. 
External horizontal input provides deflection factors of 
20 millivolts to 2 
volts/division for external sweep signal. 
The crt is a direct-view, bistable storage device having 
an 8x lodivision internal graticule. Each graticuledivision 
equals  0.25-inch.  An  enhance  mode  increases  single- 
sweep writing rate. 
The internal 1 kilohertz calibrator is connected inter- 
nally to the vertical inputs and to a front-panel connector. 
The 314 operates from a line voltage of 100, 115, 120, 
200, 230, or 240 volts ac, or from a 
+12-  or +24-volt  dc 
source. The following tables contain a detailed listing of 
the 314 characteristics. 
TABLE 
1-1 
Electrical Characteristics 
Deflection Factor 
Characteristic 
Calibrated Range 
Accuracy 
Uncalibrated Range 
(VOLTS/DIV VARIABLE) 
VERTICAL 
Performance Requirement 
Frequency Response 
Supplemental  Information 
Bandwidth 
Direct Coupled, 
4-div reference 
Capacitively Coupled, 
4-div reference 
Step Response 
Risetime, 4-division 
Step Input 
1 
mV to 10 V/Div 
Within 3% over the calibrated range 
Continuously variable between calibrated 
deflection factor settings. Extends 
highest deflection factor to at 
least 25  V/Div. 
13 steps in a 1-2-5 sequence 
REV. 
A. AUG. 
1977 
1 mV to 10 V/Div  with 
VARIABLE at CAL 
1 
mV to 10 V/Div  with 
VARIABLE at CAL 
Without Probe 
At least 
10 MHz 
10 Hz or less 
to at least 
10 MHz 
36 ns or less 
With 
P6149 Probe 
At  least 
10 MHz 
1 Hz or less to 
at least 
10 MHz 
1 mV to 10 V/Div