Performance CheckICalibration-314  Service 
DISPLAY 
Equipment Required 
1.  Digital Voltmeter 
3.  50-ohm BNC Cable 
2.  Time Mark Generator 
4.  50-ohm BNC Termination 
Control Settings 
See 
in the pullout Diagrams section 
Same  as  preset  in  POWER  SUPPLY  AND 
CALIBRATOR section, except pull POWER-INTENSITY 
switch on. 
1.  Adjust Intensity Limit 
a.  Connect DVM between TP695 and TP696 (see Fig. 
7-3  for  location,  make sure  DVM negative  lead  is  not 
connected to ground). 
b.  Turn INTENSITY control fully clockwise. 
c.  Check  for  a  DVM  reading  of  0.90 V,  i49.5 mV 
(0.8505 to 0.9495 V). 
d.  ADJUST-R702  (see  Fig.  7-3),  lntens  Limit, for 
0.90 V. 
e.  Turn INTENSITY control fully counterclockwise. 
2. 
Adjust Astigmatism 
a.  Connect  1 ms  time  markers  from  Time-Mark 
Generatorthrough 50-ohm cableand 50-ohm termination, 
to CH 1 VERT  INPUT connector. Set TIME/DIV to 1 ms. 
b.  Set DISPLAY to CH 1, CH 
1 
lnput Coupling to DC, 
and VOLTS/DIV  and Variable for 2 divisions of  vertical 
display.  Set  INTENSITY  and  TRIGGERING  LEVEL  as 
necessary to obtain display. 
c.  Set  INTENSITY  midway  between a barely visible 
trace and fully clockwise. 
d. 
ADJUST-R651,  Astig  (see  Fig. 7-4  for  location), 
and FOCUS for best definition of time marks over the full 
screen  (using CH 1 POSITION control to  position the 
display). 
3. 
Adjust Trace Rotation 
a.  Set  CH  1  lnput  Coupling  to  GND  and  observe 
horizontal trace. 
f.  Set 
TIME/DIV to 10 ms. 
b.  Position  trace  vertically  to  graticule  center  line. 
Adjust INTENSITY as necessary for thin trace. 
g.  Note voltage reading on DVM. 
c.  Check that trace is parallel to graticule center line. 
h.  Rotate INTENSITY control fully clockwise. 
i.  Adjust R706 (SN300200& up) forareadingof0.15 V 
higher  than  the  reading  noted in step  1, part g. If the 
correct  reading  cannot  be  obtained,  select  a  value 
(between  1.2 M and 2.7 M) for  R708 that  results in the 
correct reading when R706 is adjusted. 
j. 
Remove the DVM leads from TP695 and TP696. 
k.  Return TIME/DIV to 1 ms. 
d.  ADJUST-R653,  Trace  Rotation (see Fig. 7-3  for 
location), to align trace with center graticule line (position 
vertically as  necessary). 
4.  Adjust Geometry 
a.  Set CH 
1 
lnput Coupling to DC. 
b.  Turn CH 1 POSITION control to set time marker 
base line to bottom graticule line. 
REV. 
A, 
AUG. 
1975