Performance CheckICalibration-314  Service 
g.  Switch  CH  1  VOLTS/DIV  to  1 
m 
and  return 
generator  frequency  to  50 kHz.  Place  10X  attenuator 
between signal cable and termination. 
h.  Adjust generator output amplitude for 
4 
divisions of 
vertical display. 
a.  Connect Square-Wave Generator Fast-Rise, 
+ 
Out- 
put through  GR  to BNC female  adapter; 50-ohm cable; 
lox, 50-ohm attenuator; 2X, 50-ohm attenuator; and 50- 
ohm termination, to CH 1 VERT  INPUT connector. 
b.  Adjust  generator 
+ 
Transition  Amplitude  for a 
4 
division  vertical  display  on  graticule.  Adjust 
TRIGGERING LEVEL for stable display. See Fig. 5-1. 
i.  Increase  generator  frequency  until  the  display 
amplitude decreases to 2.8 divisions 
(-3 
dB point). 
c.  CHECK-risetime  (time  between  10  and  90% 
amplitude  points)  should  be 
<36 
nanoseconds 
(36 
ns 
j.  CHECK-the  generator frequency should be at least 
10 
MHz. 
k.  Switch DISPLAY to CH 
2 
and move generator signal 
to CH 
2 
VERT  INPUT connector. 
I.  Switch generator frequency to 50 kHz. Set generator 
amplitude for 
4 
divisions of vertical display. 
m.  Repeat  parts  e  through  j  for  CH 2.  Disconnect 
Constant Amplitude Signal Generator. 
8. 
Check Risetime 
Reset controls as  follows: 
DISPLAY  CH 1 
CH 
1 
VOLTSIDIV 
10 
m 
TIME/DIV  1 I.rs 
SWEEP MAG  X10 
TRIGGERING Mode  NORM 
equals  1.8 minor horizontal division). See Fig. 5-1. 
Fig. 
5-1. 
Idealized display 
of 
risetime.