EasyManuals Logo
Home>Tektronix>Test Equipment>314

Tektronix 314 User Manual

Tektronix 314
166 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
Performance CheckICalibration-314 Service
e. Switch CH 2 VOLTS/DIV to 10 m and repeat parts b
Fig. 7-4 for location of AA and AB on Vertical circuit board
through d until there is no shift while switching between (make sure DVM negative lead is not connected to
10 m and 5 m.
ground).
f. Rotate CH 2
VOLTS/DIV VARIABLE throughout its b. Set DVM range to 2 V and set CH
1
POSITION
range.
control for a zero-volt reading.
g. ADJUST4371 (see Fig. 7-4), Var Bal, until there is
c. Reconnect DVM (set to 200-volt range) between
no trace shift while rotating VARIABLE from fully upper deflection plate (terminal AB) and ground.
counterclockwise to fully clockwise.
d.
ADJUST43182 (see Fig. 7-4), DFLP DC Level, fora
h. INTERACTION-it may be necessary to repeat DVM reading of +50 V.
parts b through g because of some interaction between
Step
Atten Bal and VARIABLE controls. Return CH 2
VARIABLE to CAL.
5.
Adjust CH
2
Limit Centering
a. Set DISPLAY to CH 2.
3.
Adjust CH 1 Limit Centering
Set controls as follows:
TIME/DIV 10 ps
CH 1 VOLTS/DIV 10 m
DISPLAY CH
1
Input Coupling DC
b.
Connect DVM between upper deflection plate (ter-
minal AB) and lower deflection plate (terminal AA) on
Vertical circuit board. Set DVM to 2-volt range. See Fig. 7-4
for locations of terminals AA and AB.
c. Adjust CH 2 POSITION control for a zero-volt DVM
a. Connect Constant Amplitude Signal Generator
out-
reading,
put to CH 1 VERT INPUT connector via 50-ohm cable and
50-ohm termination.
d. Reconnect DVM (set to 200-volt range) between
upper deflection plate (terminal AB) and ground.
b. Connect test oscilloscope 1 OX probe ground lead to
314 chassis ground.
e. ADJUST-493 (see Fig. 7-4), Limit Centering, for a
DVM reading of +50 V.
c. Connect probe tip to Q150 collector. See Fig. 7-4 for
location.
6.
Adiust Add Deflection Plate DC Level
d. Set Constant Amplitude Signal Generator frequen-
a. Switch DISPLAY to ADD (both CHOP and ALT
cy to 50
kHz
and adjust output amplitude to produce a
buttons
in),
clipped sinewave display on the test oscilloscope screen
(at 10
mV/Div). Reduce signal generator output amplitude
to the point at which clipping of the
sinewave just begins.
b. Connect DVM (set to 2-volt range) between upper
deflection plate (terminal AB) and lower deflection plate
(terminal AA).
e.
ADJUST-R53 (see Fig. 7-4), Limit Centering, until
clipping on the test oscilloscope display is equal at top and
bottom.
c. Adjust either CH 1 or CH 2 POSITION control for
zero-volt DVM reading.
f. Remove probe and disconnect
sinewave generator
signal from 314.
d. Reconnect DVM (set to 200-volt range) between
upper deflection plate (terminal AB) and ground.
4. Adjust Deflection Plate DC Level
a. Connect DVM between upper deflection plate (ter-
e. ADJUST-R108 (see Fig. 7-4), Add Level,fora DVM
minal AB) and lower deflection plate (terminal AA). See
reading of i-50 V.
REV.
A,
AUG.
1975

Table of Contents

Other manuals for Tektronix 314

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 314 and is the answer not in the manual?

Tektronix 314 Specifications

General IconGeneral
BrandTektronix
Model314
CategoryTest Equipment
LanguageEnglish

Related product manuals