EasyManua.ls Logo

Tektronix 370B - Table 4-10: Display Data and CRT Readout IO

Tektronix 370B
216 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GPIB
4-38
370B User Manual
Table 4-10: Display Data and CRT Readout I/O
WAVEFORM TRANSFER COMMAND GROUP
Command Argument Link Definition
WFMpre <string> Load waveform preamble data.
WFMPRE WFID:<wfid>,ENCDG :BIN,
NR.PT:<point>,PT.FMT:XY, XMULT:<x multi>,
XZERO:0, XOFF:<x off>,XUNIT:V, YMULT:<y multi> ,
YZERO :0 , YOFF :<y off> ,YUN IT:A, BYT/NR:2,
BN.FMT:RP,BIT/NR:10,CRVCHK:CHKSMO,
LN.FMT: <format>
<wfid> ::=INDEX <num>/VERT<amp>/HORIZ <volt>/
STEP<step>/OFFSET<offset)/BGM<para>/
AUX<aux>/ACQ<acq>/VCS<percent>/TEXT <txt>"
<num> ::= Memory location number
<amp> ::= Vertical amps/div
<volt> ::= Horizontal volts/div
<step> ::= Step amplitude
<offset> ::= Step offset
<para> ::= Beta or gm
<aux> ::= Aux Supply
<acq> ::= Curve acquisition mode (AVG|NOR|ENV)
<percent> ::= Variable Collector Supply setting %
<txt> ::= Text area readout
<point> ::= curve point no. (1 - 1024)
<x multi> ::= <NR3>= horizontal scale factor
<x off> ::= <NR1>= horizontal scale offset
<y multi> ::= <NR3>= vertical scale factor
<y off> ::= <NR1>= vertical scale offset
<format> ::= VECTOR|DOT|SWEEP <num><num> ::=
number of steps + 1
WFMpre NR.PT <NR1> Sets input points number from curve command (1 to 1024)

Table of Contents