Getting Started
370B User Manual 1-9
Figure 1-3: Rackmount hole spacing
When rackmounting the 370B, take note of the following: Allow one inch
clearance above and below, and on the left and right sides of the 370B for air
circulation. Allow at least three inches of clearance between the 370B rear panel
and the rack enclosure for adequate cooling air and to provide cable clearance.
The depth of the 370B from behind the rack ears to the rear panel is480mm
(18.9in).The rack depth must be at least 559 mm(22 in) (see Figure 1–4)to meet
the rear clearance requirement.
Figure 1-4: Rackmounting Length and Clearance
The 370B is 312mm(12.25 in) high ,a multiple of 45mm(1.75 in) (the standard
rack spacing). If the 370Bis installed in a rack with standard hole spacing. and