Adjustment Procedures
3. Click the Inter
-Channel Skew... buttonintheServiceUImenutodisplay
the following setup window:
Figure 3-5: Inter-Channel Skew Calibration setup window
4. Perform the following measurements:
a. Adj
ust the horizontal position of sampling oscilloscope so that the
rising edge of the arbitrary waveform generat or Ch 1 Analog (+) Output
waveform is visible near the center of the oscilloscope screen, and then
select File Menu > Save Waveform... to save the waveform data to R1.
b. Disconnect the SMA cable from the Ch 1 Analog (+) Output, and then
connect it to the Ch 2 Analog (+) Output. Do not use a different cable for
the Ch 1 and Ch 2 connectors. Always use the same SMA cable.
c. Push CLEAR DATA and then RUN/STOP on the sampling oscilloscope
front panel control.
AWG7000B and AWG7000C Series Service Manual 3–9