Commands Listed in Alphabetical Order
Syntax
DATa:DESTinat
ion {REF< x>}
DATa:DESTination?
Related Commands
CURVe, WAVFrm
?, DATa:STARt, DATa:STOP
Arguments
REF<X> REF1 through REF8 are the reference memory locations where the
waveform wi
ll be stored. You cannot have a channel as a destination.
Examples
DATA:DESTINATION REF3 stores the incoming waveform data in Reference 3.
DATA:DESTINATION? might return DATA:DESTIN ATION REF8 as the reference
memory location that is currently selected.
DATa:ENCdg
This command sets or queries the format of the waveform data. This command
is equivalent to setting WFMOutpre:ENCdg, WFMOutpre:BN_Fmt, and
WFMOutpre:BYT_Or, a s shown in the table below. Only signed, 4-byte integers
and oating point values are supported. Changing the DATa :ENCdg parameter
will coerce WFMOutpre formatter commands accordingly. If ASCIi is selected,
the o
utput will be formatted to oating point or signed integer, based on how the
data is stored in the instrument.
Group
Wav
eform Transfer
Syntax
DATa:ENCdg { ASCIi | RIBinary | FPBinary | SRIbinary |
SFPBinary }
DATa:ENCdg?
Related Commands
WFMOutpre:ENCdg, WFMOutpre:BN_Fmt, WFMOutpre:BYT_Or
2-11 0 DSA/CSA/TDS8X00/B Series Programmer Manual