Commands Listed in Alphabetical Order
HORizontal:R
EF<x>:MAIn:SCAle? (Query O nly)
This query only command returns the scale (time per division) of the Main
timebase for the specied reference waveform. The reference waveform is
specied by x
, which can be 1 through 8.
Thehorizontalscalethatisreturnedwith this command is unique to the specied
reference w
aveform. The horizontal scale, together with the record length value,
determines the sample interval.
Group
Horizontal
Syntax
HORizontal:REF<x>:MAIn:SCA le?
Related Commands
HORizontal:REF<x>:MAIn:RECordlength?, HORizontal:REF<x>:MAIn:
RESolution?
Examples
HORIZONTAL:REF3:MAIN:SCALE ? might return
HORIZONTAL:REF3:MAIN:SCALE 2.00000000000E –009, indicating that the
Main timebase is 2 ns per division for the Reference 3 waveform.
HORizontal:REF<x>:MAIn:TOFPoint? (Query Only)
This query only command returns time of rst point of the specied reference
waveform record using the Main timebase. The time of rst point is the time
interval between the trigger point and the rst point in the speci ed Reference
Main timebase record.
Group
Horizontal
Syntax
HORizontal:REF<x>:MAIn:TOF Point?
Related Commands
HORizontal:REF<x>:MAIn:POSition?, HORizontal:REF<x>:MAIn:REFPoint?,
HORizontal:REF<x>:MAIn:SCAle?
Examples
HORIZONTAL:REF1:MAIN:TOFPO INT? might return
HORIZONTAL:REF1:MAIN:TOFPO INT 41.36300000000E-009, indicating that
the time interval between the trigger point and the rst point in the record for
the Reference 1 Main timebase record is 41.36 ns.
DSA/CSA/TDS8X00/B Series Programmer Manual 2-191