Commands Listed in Alphabetical Order
Related Commands
ACQuire:STOPA
fter:COUNt, ACQuire:STOPAfter:CONDition
Returns
NR1 is the current count value of waveforms accumulated for mask testing.
Examples
ACQUIRE:CURRENTCOUNT:MASKW FMS? might return
ACQUIRE:CURRENTCOUNT:MASKW FMS 25, indicating that currently 25
waveforms
have been acquired for mask testing.
ACQuire:DATA:CLEar (No Query Form)
This command (no query form) causes an acquisition reset and clears all acquired
data, clears the display, and, if active, resets FrameScan. Whe n a clear data
occurs, it has the following effects:
Sample-mode waveform. The current waveform data is replaced by the
waveform data of the next acquisition cycle when it is available.
Average-mode w aveform. The averag e a lgorithm mode resets such that the
next acquisition will be a straight-through copy.
Envelope-mode waveform. The envelope algorithm mode resets such that the
next acquisition will be a straight-through copy.
Waveform database. T he contents of the database will be cleared (reset
to zero) at the start of the next acquisition cycle before adding the next
acquisition.
Histogram data and statistics. The data and all statistics will be cleared
immediately.
Mask counts and statistics. Mask counts and all statistics will be cleared
immediately.
Acquisition status. The acquisition waveform count or StopAfter condition is
reset and target counts are reloaded.
Counts. Resets all counts, including number of acquired w aveforms,
acquisition and math average counts, conditional stop counts, and FrameScan
counts (if any of these are active).
Measurement statistics. Measurement statistics are reset.
Group
Acquisition
Syntax
ACQuire:DATA:CLEar
Examples
ACQUIRE:DATA:CLEAR causes an acquisition reset and clears all acquired data.
DSA/CSA/TDS8X00/B Series Programmer Manual 2-55