Commands Listed in Alphabetical Order
Examples
ACQUIRE:NUMAV
G10
specifies that an averaged waveform will show the result
of combining 10 separately acquired waveforms.
ACQUIRE:NUMAVG? might return ACQUIRE:NUMAVG 75, indicating that there are
75 acquisitions specified for averaging.
ACQuire:SAVEFile:SAVEScreen
This command sets or queries the file to which a screen is saved at the completion
of a StopAfter condition (when the StopAfter action is set to SAVEScreen). See
the ACQuire:STOPAfter:ACTion command. The screen is saved in the specified
file in .BMP format. This is equivalent to selecting Print Screen to File in the Stop
Action section of the Acquisition Setup dialog box and entering a file name.
Group
Acquisition
Syntax
ACQuire:SAVEFile:SAVEScre en <Qstring>
ACQuire:SAVEFile:SAVEScre en?
Related Commands
ACQuire:STOPAfter:ACTion, ACQuire:STOPAfter:CONDition,
ACQuire:STOPAfter:MODe
Arguments
<Qstring> defines the file name and path. Input the file path using the form
<drive>/<dir>/<filename>. The <drive> and one or more <dir> arguments are
optional. If you do not specify them, the instrument will copy the file into the
current directory. The <filename> can be a Windows long file name. Do not
use wild card characters.
Examples
ACQUIRE:SAVEFILE:SAVESCRE EN "Wavetest.bmp" specifies the file
Wavetest.bmp (in the current directory) to which the screen will be saved after a
StopAfter SAVEScreen condition is met.
ACQUIRE:SAVEFILE:SAVESCRE EN? might return
ACQUIRE:SAVEFILE:SAVESCRE EN "TESTFILE.BMP", indicating that the
screen will be saved in TESTFILE.BMP in the current directory when the
StopAfter action is enabled.
ACQuire:SAVEFile:SAVEWfm
This command sets or queries the file to which waveform data is saved at
the completion of a StopAfter condition (when the StopAfter action is set to
SAVEWaveform). See the ACQuire:STOPAfter:ACTion command. This is
equivalent to selecting Save All Waveforms in the Stop Action section of the
2-58 DSA/CSA/TDS8X00/B Series Programmer Manual