Section
10: Capturing and analyzing waveforms Model DMM7510 7½ Digit Multimeter
User's Manual
10-12 DMM7510-900-01 Rev. C / September 2019
Inductor current measurement device connections
You can use either the front or rear inputs. Ensure that the front-panel TERMINALS switch is set
correctly (F displayed for front; R displayed for rear). You must use either the front terminals or rear
terminals. You cannot mix connections.
Both front-panel and rear-panel connections are safety banana jacks. You can make these
connections with two insulated banana cables (for example, the leads included with the Keithley
Instruments Model 1756 General Purpose Test Lead Kit).
Connect the test leads as shown in the following figure.
Figure 43: Inductor current linearity test connections
Run the test using the front panel
Running the inductor current linearity test from the front panel:
1. Press the POWER button on the front panel to turn on the instrument.
2. On the FUNCTIONS swipe screen, select Digi I to select the digitize current function.
3. Swipe to the SETTINGS swipe screen.
4. Set the Sample Rate to 500,000.
5. Set the Count to 50.
6. Press the MENU key.
7. Select Graph to view the graph.

Table of Contents

Other manuals for Tektronix Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley DMM7510 and is the answer not in the manual?

Tektronix Keithley DMM7510 Specifications

General IconGeneral
BrandTektronix
ModelKeithley DMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals