Commands Listed in Alphabetical Order
condition is se
tto
DATa, and the qualifier is set to INrange or OUT range,
this command specifies the lower limit of the range. (Use the command
TRIGger:A:BUS:B<x>:ETHERnet:DATa:HIVALue to set the upper limit of the
range.) The default is all X’s (don’t care).
B<x> is the bus number B1-B4.
Conditions
This command requires a DPO4ENET application module for DPO/MSO4000B
and MDO4000/B/C models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:ETHERnet:DATa:VALue <QString>
TRIGger:A:BUS:B<x>:ETHERnet:DATa:VALue ?
Related Commands
TRIGger:A:BUS:B<x>:ETHERnet:CONDition
TRIGger:A:BUS:B<x>:ETHERnet:QUALifier
TRIGger:A:BUS:B<x>:ETHERnet:DATa:OFFSet
TRIGger:A:BUS:B<x>:ETHERnet:DATa:SIZe
TRIGger:A:BUS:B<x>:ETHERnet:DATa:HIVALue
Arguments
QString is a quoted string where the allowable characters are 0, 1, and X.
The allowable number of characters depends on the setting for size (using
TRIGger:A:BUS:B<x>:ETHERnet:DATa:SIZe). The bits specified in the
quoted string replace the least significant bits, leaving any unspecified upper
bits unchanged.
Examples
TRIGger:A:BUS:B1:ETHERnet:DATa:VALue “00001000” sets the binary
data to trigger on to 00001000, assuming the qualifier is set to
LESSthan,
MOREthan, EQual, UN EQual, LESSEQual
or MOREEQual, and DATa:SIZe
is set to 1 byte.
TRIGger:A:BUS:B1:ETHERnet:DATa:VALue “00001000” sets the lower
limit of the range to 00001000, assuming the qualifier is set to
INrange or
OUTrange,andDATa:SIZ e is set to 1 byte.
TRIGger:A:BUS:B<x>:ETHERnet:FRAMETYPe
This command specifies the Ethernet frame type: either Basic or QTag (IEEE
802.1Q, or VLAN tagging). The default is Basic.
B<x> is the bus number B1-B4.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-797