Commands Listed in Alphabetical Order
Related Commands
TRIGger:A:BUS
:B<x>:ETHERnet:CONDition
TRIGger:A:BUS:B<x>:ETHERnet:QUALifier
TRIGger:A:BUS:B<x>:ETHERnet:MAC{:LENgth|:TYPe}:VALue
Arguments
QString is a quoted string of up toe 16 characters where the allowable characters
are 0, 1, and X. The bits specified in the quoted string replace the least significant
bits, leaving any unspecified upper bits unchanged.
Examples
TRIGger:A:BUS:B1:ETHERnet:MAC{:LENgth| :TYPe}:HIVALue
“XXXXXXXX00001000” sets the upper limit of the range to the hexadecimal
value XX08 (when the trigger condition is set to
MACLENgth, and the qualifier is
set to
INrange or OUTrange).
TRIGGER:A:BUS:B1:ETHERNET:MAC{:LENGTH| :TYPE}:HIVALUE?
might
return “XXXXXXXX00001000”.
TRIGger:A:BUS:B<x>:ETHERnet:MAC{:LENgth|:TYPe}:VALue
When the Ethernet trigger condition is set to MACLEN gth , and the qualifier is set
to
LESSthan, MOREthan, E Qual, UNEQual, LESSEQual or MOREEQual,
this command specifies the 16–bit value to trigger on. When the qualifier is set
to
INrange or OUTrange, this command specifies the lower limit of the range.
(Use the command TRIGger:A:BUS:B<x>:ETHERnet:MAC{:LENgth|:TYPe}:
HIVALue to set the upper limit of the range.) The default is all X’s (don’t care).
B<x> is the bus number B1-B4.
Conditions
This command requires a DPO4ENET application module for DPO/MSO4000B
and MDO4000/B/C models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:ETHERnet:MAC{:LENgt h|:TYPe}:VALue
<QString>
TRIGger:A:BUS:B<x>:ETHERnet:MAC{:LENgt h|:TYPe}:VALue?
Related Commands
TRIGger:A:BUS:B<x>:ETHERnet:CONDition
TRIGger:A:BUS:B<x>:ETHERnet:QUALifier
TRIGger:A:BUS:B<x>:ETHERnet:MAC{:LENgth|:TYPe}:HIVALue
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-803