Commands Listed in Alphabetical Order
CALibrate:RE
Sults:FACtory:RF? (Query Only)
This query returns the factory calibration status for the RF portion of the
instrument, if present.
This query is synonymous with
CALibrate:FACtory:STATus:RF?
Group
Calibration and Diagnostic
Syntax
CALibrate:RESults:FACtory:RF?
Returns
PASS - factory calibration has passed
INIT - factory calibration is unadjusted
RUNNING - factory calibration is in progress
Examples
CALIBRATE:RESULTS:FACTORY:RF? might return RUNNING, indicating that the
factory calibration for the RF portion of the instrument is in progress.
CALibrate:RESults:FACtory:SCOPE? (Query Only)
This query returns the factory calibration status for the oscilloscope (doesn't
include RF or AFG) of the instrument.
This query is synonymous with the following query:
:CALibrateFACtory:STATus:SCOPE?
Group
Calibration and Diagnostic
Syntax
CALibrate:RESults:FACtory:SCOPE?
Returns
This query will return one of the following:
INIT indicates the oscilloscope portion of the instrument has not been calibrated.
PASS indicates the oscilloscope internal calibration completed successfully.
RUNNING indicates the oscilloscope internal calibration is currently running.
Examples
CAL:RESULTS:SPC:SCOPE? might return I NIT indicating that the oscilloscope
portion of the instrument has not been calibrated.
2-216 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual