Commands Listed in Alphabetical Order
Syntax
CH<x>?
CH<x>:AMPSVIAVOLTs:ENAble
This command specifies the state of the amps via volts feature for the specified
channel. This feature supports measuring current via the voltage drop across a
resistor (1–4).
Group
Vertical
Syntax
CH<x>:AMPSVIAVOLTs:ENAble {<NR1>|OFF|ON}
CH<x>:AMPSVIAVOLTs:ENAble?
Arguments
OFF sets the amps via volts function for channel <x> to off.
ON sets the amps via volts function for channel <x> to on.
<NR1> = 0 sets the amps via volts function to off. Any other value sets the
functiontoon.
Examples
CH1:AMSVIAVOLTS:ENABLE ON turns on the amps via volts feature for
channel 1.
CH<x>:AMPSVIAVOLTs:FACtor
This command specifies the amps via volts factor for the specified channel (1–4).
Group
Vertical
Syntax
CH<x>:AMPSVIAVOLTs:FACtor <NR3>
CH<x>:AMPSVIAVOLTs:FACtor?
Arguments
<NR3> is a double-precision ASCII string that represents the amps via volts factor.
Examples
CH1:AMPSVIAVOLTS:FACTOR 15.5 sets the amps via volts factor for channel 1
to 15.5.