Commands Listed in Alphabetical Order
Syntax
CH<x>:PRObe:C
OMMAND <Q String>, <QString>
Arguments
<QString> are quoted strings specifying the probe command and value to set in
the probe attached to the specified channel.
Examples
CH1:PROBE:COMMAND “MODE”, “4–4V1MHz” sets a Tektronix VPI-DPG probe
to the 4-4V1MHz mode.
CH1:PROBE:COMMAND “OUTPUT”, “ON turns the output of a Tektronix
VPI-DPG probe on.
CH1:PROBE:COMMAND?“MODE” might return CH1:PROBE:COMMAND
“MODE”,“4-4V1MHZ”
.
CH<x>:PRObe:DEGAUss (No Query Form)
This command starts a degauss auto-zero cycle on a TekVPI current probe
attached to the input channel specified by <x>, where x is the channel number.
Group
Vertical
Syntax
CH<x>:PRObe:DEGAUss EXECute
Arguments
EXECute initiates the degauss operation.
CH<x>:PRObe:DEGAUss:STATE? (Query Only)
This command returns the state of the probe degauss for the channel specifiedby
<x>, where is x is the channel number.
NOTE. This command w ill return
PASSED
for probes that do not support degauss
operations.
Group
Vertical
Syntax
CH<x>:PRObe:DEGAUss:STATE?
Returns
NEEDED indicates the prob
e should be degaussed before taking measurements.
2-228 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual