Commands Listed in Alphabetical Order
DVM:MEASUrem
ent:HIStory:AVErage? (Query Only)
Returns the average readout value for the Digital Voltmeter function over the
history period.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:MEASUrement:HIStory:AVErage?
Examples
DVM:MEASU:HIS:AVE? might return 100.0000E+3, which represents the
average DVM readout value over the history period.
DVM:MEASUrement:HIStory:MAXimum? (Query Only)
Returns the maximum readout value for the DVM function over the history period.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:MEASUrement:HIStory:MAXimum?
Examples
DVM:MEASU:HIS:MAX? might return 100.0000E+3, which represents the
maximum readout value for the DVM function over the history period.
DVM:MEASUrement:HIStory:MINImum? (Query Only)
Returns the minimum readout value for the DVM function over the history period.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
2-298 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual