Commands Listed in Alphabetical Order
RF_AMPlitude|
RF_FREQuency|RF_PHASe
RF time domain traces
(MDO4000/B/C models only.)
The next 7 <NR3
> arguments are time mark parameters as follows:
Start start value in percentage of record
End end value in percentage of record
Focus focus value in percentage of record
X value horizontal value of the focus point in the mark’s waveform units
(This is ignored in the command form use any value.)
Y value vertical value of the focus point in the mark’s waveform units (This
is ignored in the command form use any value.)
Delta X value always 0, since there is no time-domain reference mark. (This
is ignored in the command form use any value.)
Delta Y value always 0, since there is no time-domain reference mark. (This
is ignored in the command form use any value.)
Returns
If no user time marks have been set, the query returns
NONe,0.0E0,0.0E0,0.0E0,0.0E0,0.0E0,0.0E0,0.0E0.
Examples
MARK:USERLIST
CH1,62.5000,62.5000,62.5000,0E+0,0E+0,0.0E+0,0.0E+0;USERLIST
CH1,79.5000,79.5000,79.5000,0E+0,0E+0,0.0E+0,0.0E+0 sets two
marks on the Channel 1 waveform.
MARK:USERLIST? might return
CH1,62.5000,62.5000,62.5000,5.0000E-6,infE-315,0.0E+0,0.0E+0;USERLIST
CH1,79.5000,79.5000,79.5000,11.8000E-6,infE-315,0.0E+0,0.0E+0, indicating
t
hat two marks were set on the Channel 1 waveform.
MARKER:MANual
This command switches on or off the manual markers a and b that are available
for frequency domain traces.
Two manual markers are available for measuring non-peak areas of interest.The
absolute measurements are in dBm; the relative measurements (relative to the
Reference Marker) are in dBc (dB relative to the carrier).
When the manual markers are turned off, and the peak markers are turned on, the
Reference Marker is placed on the highest amplitude peak. With manual markers
on, the Reference Marker is placed at the a manual marker. The manual markers
use the units specified with the command RF:UNIts.
2-358 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Table of Contents

Other manuals for Tektronix MDO4000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
BrandTektronix
ModelMDO4000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals