Commands Listed in Alphabetical Order
Aseriesofexam
ples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:DISplay {ON|OFF|1|0}
MASK:DISplay?
Related Commands
MASK:TEMPLate:CREATEmask
Arguments
ON shows the mask on the display. This is the default value.
OFF removes the mask from the display.
0 removes the mask from the display;
1 shows the mask on the display.
Examples
MASK:DISplay ON shows the mask on the display.
MASK:DISplay? might return 0, indicating that the mask is not being shown
on the display.
MASK:LOCk
This command locks the mask to the source waveform so that any changes made
to the horizontal and/or vertical scale settings of the waveform will redraw the
mask segments in proportion. This feature is useful for expanding the horizontal
and/or vertical settings in order to zoom in on waveforms and masks, and visually
examine violation areas in more detail.
A series of examples showing how to use mask commands for t
ypical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
2-376 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual