EasyManua.ls Logo

Tektronix MDO4000B Series

Tektronix MDO4000B Series
1158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands Listed in Alphabetical Order
:MASK:TEMPLat
e:TOLerance:HORizontal 0.4
:MASK:TEMPLate:CREATEmask
Notice that the horizontal gap in the mask is twice as tall.
MASK:TEMPLate:TOLerance:VERTical
This comman
d specifies the vertical limit (tolerance) for a template mask to be
used for limit testing. This indicates how far vertically from the templatesource
to create the mask.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TEMPLate:TOLerance:VERTical <NR3>
MASK:TEMPLate:TOLerance:VERTical?
Related Commands
MASK:TEMPLate:CREATEmask, MASK:TEMPLate:TOLerance:HORizontal,
SAVe:MASK
Arguments
<NR3> is a floating point value that specifies the vertical limit (tolerence) in
milli-divisions. Range: 0 to 1000 mDiv, resolution: 1 mDiv, default: 40mDiv.
This value cannot be negative.
Examples
If you create a limit mask after a default setup, then query:
:MASK:TEMPLate:TOLerance:VERTical?
you get the response: 200.000E
-3
Now send the commands:
:MASK:TEMPLate:TOLerance:VERTical 0.4
:MASK:TEMPLate:CREATEmask
Notice that the vertical gap in the mask is twice as tall.
2-384 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual