Commands Listed in Alphabetical Order
:MASK:TEMPLat
e:TOLerance:HORizontal 0.4
:MASK:TEMPLate:CREATEmask
Notice that the horizontal gap in the mask is twice as tall.
MASK:TEMPLate:TOLerance:VERTical
This comman
d specifies the vertical limit (tolerance) for a template mask to be
used for limit testing. This indicates how far vertically from the templatesource
to create the mask.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TEMPLate:TOLerance:VERTical <NR3>
MASK:TEMPLate:TOLerance:VERTical?
Related Commands
MASK:TEMPLate:CREATEmask, MASK:TEMPLate:TOLerance:HORizontal,
SAVe:MASK
Arguments
<NR3> is a floating point value that specifies the vertical limit (tolerence) in
milli-divisions. Range: 0 to 1000 mDiv, resolution: 1 mDiv, default: 40mDiv.
This value cannot be negative.
Examples
If you create a limit mask after a default setup, then query:
:MASK:TEMPLate:TOLerance:VERTical?
you get the response: 200.000E
-3
Now send the commands:
:MASK:TEMPLate:TOLerance:VERTical 0.4
:MASK:TEMPLate:CREATEmask
Notice that the vertical gap in the mask is twice as tall.
2-384 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual