Commands Listed in Alphabetical Order
Group
Mask
Syntax
MASK:USER:AMPLitude <NR3>
MASK:USER:AMPLitude?
Related Commands
MASK:CUSTom,
SAVe:MASK
Arguments
<NR3> is a floating point number that determines the nominal pulse amplitude, in
volts, of a user-defined custom mask.
Examples
MASK:USER:AMPLITUDE 2.4 sets the nominal pulse amplitude of a user-defined
custom mask to 2.4 volts.
MASK:USER:AMPLITUDE? might return
:MASK:USER:AMPLITUDE 100.0000E–3, indicating that the user mask nominal
pulse amplitude is set to 0.1 volts.
MASK:USER:HSCAle
This command specifies the nominal timing resolution, in time/division, tobe
used to draw a user-defined custom mask pulse shape.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
andanMDO3LMTapplicati
on module for MDO3000 series models.
Group
Mask
Syntax
MASK:USER:HSCAle <NR3>
MASK:USER:HSCAle?
Related Commands
MASK:CUSTom,
SAVe:MASK
2-398 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual