Commands Listed in Alphabetical Order
Group
Mask
Syntax
MASK:USER:VPOS <NR3>
MASK:USER:VPOS?
Related Commands
MASK:CUSTom,
SAVe:MASK
Arguments
<NR3> is a floating point number that sets the nominal vertical position value
in divisions.
Examples
MASK:USER:VPOS -0.8 sets the user mask nominal vertical position value
to -0.8 divisions.
MASK:USER:VPOS? might return :MASK:US ER:VPOS 0.0000, indicating that
the nominal vertical position value of the displayed mask is set to 0 divisions.
MASK:USER:VSCAle
This command specifies the nominal value, in volts per division, to be used to
vertically scale the input channels for a user-defined custom mask.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:USER:VSCAle <NR3>
MASK:USER:VSCAle?
Related Commands
MASK:CUSTom,
SAVe:MASK
Arguments
<NR3> is a floating point number that sets the nominal vertical scale value for the
input channels for a user-defined custom mask.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-405