Commands Listed in Alphabetical Order
Examples
POWER:HARMONI
CS:DISPLAY:TYPE GRAph
sets the display type to graphical.
POWER:HARMONICS:DISPLAY:TYPE? might return
:POWER:HARMO
NICS:DISPLAY:TYPE TAB
indicating that the display type is
set to tabular.
POWer:HARMonics:FREQRef
This comman
d specifies the frequency reference used when the harmonic standard
is None.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:FREQRef
{VOLTage|CURRent|HARMSOURce|FIXEDFREQu ency}
POWer:HARMonics:FREQRef?
Arguments
VOLTage to use a voltage waveform as the frequency reference.
CURRent to use a current waveform as the frequency reference.
HARMSOURce to use a harmonic source waveform as the frequency reference.
FIXEDFREQuency to use a fixed frequency value instead of a waveform for the
frequency reference.
Examples
POWER:HARMONICS:FREQREF VOLTage sets voltage source as the frequency
reference.
POWER:HARMONICS:FREQREF? might return :POWER:HARMONICS:FREQREF
FIXEDFREQ
indicating that the frequency reference is a fixed value.
POWer:HARMonics:FREQRef:FIXEDFREQValue
This command specifies the fr
equency value when the :FREQRef selection is
FIXEDFREQuency.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
2-468 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual