Commands Listed in Alphabetical Order
POWer:HARMon
ics:RESults:HAR<1-400>:PHASe? (Query Only)
Returns the phase of the harmonic in degrees. The phase is measured relative
to the zero-crossing of the reference waveform. When there is no reference
waveform, th
e phase is relative to the fundamental component.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3P
WR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:HAR<1-400>:PHA Se?
Examples
POWER:
HARMONICS:RESULTS:HAR400:PHASE?
might return
:POWER:HARMONICS:RESULTS:HAR400:PHASE 0 indicating that the harmonic
phaseissetto0indegrees.
POWer:HARMonics:RESults:HAR<1-400>:RMS:ABSolute? (Query Only)
Returns the RMS magnitude of the harmonic in absolute units.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:HAR<1-400>:RMS :ABSolute?
Examples
POWER:HARMONICS:RESULTS:HAR400:RMS:ABS OLUTE? might return
:POWER:HARMONICS:RESULTS:HAR400:RMS:AB SOLUTE 1.0 indicating that
the RMS magnitude of the harmonic is set to 1.0 A.
POWer:HARMonics:RESults:HAR<1-400>:RMS:PERCent? (Query Only)
Returns the RMS magnitude of the harmonic expressed as a percentage of the
fundamental.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
2-478 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual