Commands Listed in Alphabetical Order
Syntax
POWer:QUALity
:TRUEpwr?
Examples
POWER:QUALITY:TRUEPWR? might return :POWER:QUALITY:TRU Epwr
1000 W
indicating that the true power value is set to 1000 W.
POWer:QUALity:VCRESTfactor? (Query Only)
This query returns the measurement for the voltage crest factor.
Conditions
This command requires the optional power analysis application module
DPO4PWR
Group
Power
Syntax
POWer:QUALity:VCRESTfactor?
Related Commands
POWer:QUALity:DISplay:ICRESTfactor,
APPLication:TYPe
POWer:TYPe
Returns
Floating point number.
Examples
POWer:QUALity:VCRESTfactor? might return 1.4, indicati
ng that the crest
factor is 1.4.
POWer:QUALity:VCRESTfactor? might return 9.9100E+37,
indicating that the
voltage crest factor measurement has not been computed.
POWer:QUALity:VRMS? (Query Only)
Returns the RMS voltage measurement.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
2-498 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual