Commands Listed in Alphabetical Order
use of wildcard
characters in filenames is not supported. Filename extensions are
not required, but highly recommended.
Examples
RECALL:SETUP FACTORY recalls (and makes current) the oscilloscope setup
to its factory defaults.
RECALL:SETUP 2 recalls the oscilloscope setup from setup storage location 2.
RECALL:SETUP "TEK00000.SET" recalls the setup from the file TEK00000.SET
in the current working directory.
RECAll:SETUp:DEMO<x> (No Query Form)
This command recalls one of up to 22 built-in demonstration setups of oscilloscope
functionality. <x> varies by model.
Conditions
MDO only.
Group
Save and Recall
Syntax
RECAll:SETUp:DEMO<x>
Examples
RECAll:SETUp:DEMO4 recalls the Demonstration #4 setup.
RECAll:WAVEform (No Query Form)
This command (no query form) recalls a stored waveform to a reference memory
location, and, for instruments with the arbitrary waveform feature, to arbitrary
waveform edit memory (EMEM). Only the first waveform in a .CSV file is
recalled for multiple waveform .CSV files. Recall of digital waveforms (D0
through D15) is not supported.
NOTE. For MDO3000 series models, a saved ARB waveform cannot be recalled
toareferencelocation. ItcanonlyberecalledtoanEMEMlocation.However,
an AFG waveform can be recalled into either a reference or EMEM location.
Group
Save and Recall
Syntax
RECAll:WAVEform <Source>, <Destination>
2-540 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual