Commands Listed in Alphabetical Order
TRIGger:A:BU
S:B<x>:ETHERnet:TCPHeader:SOUrceport:VALue
When the Ethernet trigger condition is set to TCPHeader, this command specifies
the 16–bit source port address that is to be used in the trigger (along with the
destination
port address, the sequence number and the acknowledgement number).
The default is all X’s (don’t care).
B<x> is the bus number B1-B4.
Conditions
This command requires a DPO4ENET application module for DPO/MSO4000B
and MDO4000/B/C models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:ETHERnet:TCPHeader: SOUrceport:VALue
<QString>
TRIGger:A:BUS:B<x>:ETHERnet:TCPHeader: SOUrceport:VALue?
Related Commands
TRIGger:A:BUS:B<x>:ETHERnet:CONDition
TRIGger:A:BUS:B<x>:ETHERnet:TCPHeader:ACKnum:VALue
TRIGger:A:BUS:B<x>:ETHERnet:TCPHeader:SEQnum:VALue
TRIGger:A:BUS:B<x>:ETHERnet:TCPHeader:DESTinationport:VALue
Arguments
QString is a quoted string of up to 16 characters where the allowable characters
are 0, 1, and X. The bits specified in the quoted string replace the least significant
bits, leaving any unspecified upper bits unchanged.
Examples
TRIGger:A:BUS:B1:ETHERnet:TCPHeader:SO Urceport:VALue
“XXXX000010100110” would set the source port address that is to be used in the
trigger to hexadecimal X0A6.
TRIGger:A:BUS:B1:ETHERnet:TCPHeader:SO Urceport:VALue? might
return “XXXXX01001010110”.
TRIGger:A:BUS:B<x>:FLEXray:CONDition
This command specifies the condition to use when triggering on a FlexRay bus
signal (start of frame, frame t
ype, ID, cycle count, header, data, ID and data,
EOF, error).
B<x> is the bus number B1-B4, except for MDO3000, which has
two buses B1–B2..
2-808 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual