Commands Listed in Alphabetical Order
TRIGger:A:BU
S:B<x>:LIN:DATa:HIVALue
This command specifies the binary data string to be used for LIN trigger if trigger
condition is ID or IDANDDATA.
B<x>
Conditions
This command requires a DPO4AUTO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AUTO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:LIN:DATa:HIVALue <QString>
TRIGger:A:BUS:B<x>:LIN:DATa:HIVALue?
Arguments
<QString> is a quoted string that is the binary data string used for LIN trigger if
the trigger condition is ID or IDANDDATA.
Examples
TRIGGER:A:BUS:B1:LIN:DATA:HIVALUE "11001010" sets the high value
to
11001010.
TRIGGER:A:BUS:B1:LIN:DATA:HIVALUE? might return
TRIGGER:A:BUS:B1:LIN:DATA:HIVALUE "XXXXXXXX"indi
cating
the high value is don't care.
TRIGger:A:BUS:B<x>:LIN:DATa:QUALier
This command specifies the LIN data qualifier. This only applies if the trigger
condition is IDANDDATA or DATA.
B<x>
Conditions
This command requires a DPO4AUTO application module for DPO/MSO4000B
and MDO4000/B/C models or
MDO3AUTO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:LIN:DATa:QUALifier
{LESSthan|MOREthan|EQual|UNEQual|LESSE Qual|MOREEQual|
INrange|OUTrange}
TRIGger:A:BUS:B<x>:LIN:DATa:QUALifier?
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-827