Commands Listed in Alphabetical Order
Arguments
<NR3> is a floati
ng point number that specifies the clock level, in volts.
TTL specifies a preset TTL high level of 1.4 V.
Examples
TRIGGER:A:SETHOLD:CLOCK:THRESHOLD TTL specifies the preset TTL
value
of 1.4 V as the clock threshold for the setup and hold trigger.
TRIGGER:A:SETHOLD:CLOCK:THRESHOLD? might return
:TRIGGER:A:LOGIC:SETHOLD:CLOCK:THRESHO LD 1.2000E+00
indicating that the clock threshold for the setup and hold trigger is 1.2 V.
TRIGger:A:SETHold:DATa? (Query Only)
Returns the voltage threshold and data source for the setup and hold trigger.
Group
Tr
igger
Syntax
TRIGger:A:SETHold:DATa?
Related Commands
TRIGger:A:SETHold:CLOCk?
Examples
TRIGGER:A:SETHOLD:DATA? might return
:TRIGGER:A:SETHOLD:DATA:SOURCE CH2;THRESHOLD 80.0000E-3
TRIGger:A:SETHold:DATa:SOUrce
This command specifies the data source or sources for setup and hold triggering.
You can specify any combination of sources as long as none of them are also
the clock source.
For MSO/MDO4000/B models, (and MDO3000 and MDO4000C models
with the MDO3MSO or MDO4MSO option installed), you can specify any
combination of CH1-CH4 and D0-D15 as the data sources. For MSO/DPO 4000B
series models, MDO4000C series models without an RF option, and 2-channel
MDO3000 models, you can also use Aux Input.
Data sources for DPO models may be one of CH1-CH4 or the Aux Input.
Group
Trigger
Syntax
TRIGger:A:SETHold:DATa:SOUrce {CH1|CH2|CH3|CH4|
D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D11| D12|D13|D14|D15| AUX}
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-901