Commands Listed in Alphabetical Order
TRIGGER:A:SET
HOLD:DATA:THRESHOLD?
might return
:TRIGGER:A:SETHOLD:DATA:THRESHOLD 1.2000E+00 indicating
that 1.2 V is the current data voltage level for the setup and hold trigger.
TRIGger:A:S
ETHold:HOLDTime
This command specifies the hold time for setup and hold violation triggering.
Group
Trigger
Syntax
TRIGger:A:SETHold:HOLDTime <NR3>
TRIGger:A:SETHold:HOLDTime?
Arguments
<NR3> is a floating point number that specifies the hold time setting, in seconds.
Positive values for hold time occur after the clock edge. Negative values oc
cur
before the clock edge.
Examples
TRIGGER:A:SETHOLD:HOLDTIME 3.0E-3 sets the hold time for the setup and
hold trigger to 3 ms.
TRIGGER:A:SETHOLD:HOLDTIME? might return
:TRIGGER:A:SETHOLD:HOLDTIME 2.0000E-09 indicating that the current
hold time for the setup and hold trigger is 2 ns.
TRIGger:A:SETHold:SETTime
This command specifies the setup time for setup and hold violation triggering.
Group
Trigger
Syntax
TRIGger:A:SETHold:SETTime <NR3>
TRIGger:A:SETHold:SETTime?
Arguments
<NR3> is a floating point number that specifies the setup time for setup and hold
viol
ation triggering.
Examples
TRIGGER:A:SETHOLD:SETTIME 3.0E-6 specifies that the current setup time for
setup and hold trigger is 3 µs.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-903