Performance Verication
10. Enter the mean v
alue as the test result in the test record.
11. Repeat steps 8 through 10 for each vertical scale setting in the test record.
12. Push the channel 1 b utton and then repeat steps 4 thorough 11 for each
bandwidth setting.
13. Repeat the tests at 1 M impedance as follows:
a. Push the front-panel channel 1 button.
b. Set the Termination (input impedance) to 1M .
c. Repeat steps 4 through 12.
14. Repeat th
e procedure for all remaining channels as follows:
a. Deselect the channel that you already tested.
b. Move the 50 terminator to the next channel input to be tested.
c. Starting from step 3, repeat the procedure for each channel.