WWXtX General Operating Instructions
This section describes the controls on the front and rear panels of the PS280
or PS283 Laboratory DC Power Supply.
Front Panel
Figure
2-1
shows the front panel of the instrument.
LABORATORY
DC POWER SUPPLY
//// TWCmniX
PS 280
CC POWER
SUPPLY/
Figure
2-1:
PS280 or PS283 Front Panel
1.
The power switch turns on the instrument when pressed. When you press
it again, the instrument is turned off.
2.
The switches labeled TRACKING determine the test mode of the instru-
ment. The PS280 or PS283 features two tracking modes: series and
parallel.
If both push-button switches are disengaged (out), the two
vari-
able power supplies operate independently. If the left switch is pushed in,
the instrument operates in series mode. If both switches are pushed in, the
instrument operates in parallel mode.
In series mode, the master power supply controls the voltage for both
power supplies, which can then range from 0 to 60 V.
In parallel mode, the master power supply controls both the voltage and
the current for both power supplies. The current can then range from 0 to
4A(0to2AforthePS283).
PS2S0 & PS283 Laboratory DC Power Supply Module-Level Service Manual
REV AUG
1991

Table of Contents

Other manuals for Tektronix PS280

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix PS280 and is the answer not in the manual?

Tektronix PS280 Specifications

General IconGeneral
BrandTektronix
ModelPS280
CategoryPower Supply
LanguageEnglish

Related product manuals