Preparation
Equipment Required
The test equipment required for both the performance verification procedures
and the adjustment procedures for the Laboratory DC Power Supply are listed
Item Number and
Description
1 Adjustment Tool
2 Multimeter, Digital
3 Resistive Load,
Vari-
4 Oscilloscope
Table
4-1:
Test
Minimum Requirements
Less than % inch
diameter and over 2
inches long
35
V,
4 A, ±0.1%
Accuracy
0 to 2 n, 30 Watts
50 MHz, 3% vertical
deflection accuracy
Equipment
Example
Tektronix part number
003-0675-00
TEKTRONIX DM252
Tektronix part number
311-2494-00
TEKTRONIX 2225
Purpose
Adjustments
Adjustments
Adjustments
Performance Verification
5 Fixed Resistive
Loads Set
20 a, 30 Watts
10 a, 30 Watts
2 fl, 5 Watts
1 a, 30 Watts
7 a, 140 Watts
0.5 a, 5 Watts
15X1,70 Watts
30 a, 140 Watts
7.511,140
Watts
1.7 n, 20 Watts
30 a, 40 Watts
60 a, 70 Watts
Tektronix part number
118-8751-00
Performance Verification
6 Mechanical Switch 60 Watts
Tektronix part number
260-2570-00
Performance Verification
REV
AUG
1991
Performance Verification