Appendix A: Specifications
TDS 200-Series Digital Oscilloscope User Manual
97
Specifications (Cont.)
Trigger
Trigger Sensitivity, Coupling Sensitivity
Edge Trigger Type
DC
CH 1, CH 2,
CH 3 & CH 4
1 div from DC to 10 MHz*,
1.5 div from 10 MHz* to Full
TDS 210 and TDS 220 only
EXT 100 mV from DC to 10 MHz*,
150 mV from 10 MHz* to Full
EXT/5 500 mV from DC to 10 MHz*,
750 mV from 10 MHz* to Full
Trigger Sensitivity, Coupling Sensitivity
Edge Trigger Type,
AC
Same as DC at 50 Hz and above
typical
NOISE REJ Reduces the DC-coupled trigger sensitivity by
2 times for > 10 mv/div to 5 V/div
HF REJ Same as the DC-coupled limit from DC to
7 kHz, attenuates signals above 80 kHz
LF REJ Same as the DC-coupled limits for frequencies
above 300 kHz, attenuates signals below
300 kHz
Trigger Level Range Source Range
Internal ±8 divisions from center of screen
TDS 210 and TDS 220 only
EXT ±1.6 V
EXT/5 ±8 V
* Bandwidth is not valid for the P2100 probe when the switch is set to 1X.

Table of Contents

Other manuals for Tektronix TDS 200-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 200-Series and is the answer not in the manual?

Tektronix TDS 200-Series Specifications

General IconGeneral
BrandTektronix
ModelTDS 200-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals