Command Groups
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−33
3. Use the DATa:WIDth command to specify the number of bytes
per data point.
4. Use the DATa:STARt and DATa:STOP commands to specify the
portion of the waveform that you want to transfer.
5. Use the WFMPRe? command to transfer waveform preamble
information.
6. Use the CURVe? command to transfer waveform data.
To the Oscilloscope. To transfer waveform data to an oscilloscope
waveform storage location, follow these steps:
1. Use the DATa:DESTination command to specify the stored
waveform location.
2. Use the DATa:ENCdg command to specify the waveform data
format.
3. Use the DATa:WIDth command to specify the number of bytes
per data point.
4. Use the DATa:STARt command to specify the first data point in
the waveform record.
5. Use the WFMPRe command to transfer waveform preamble
information.
6. Use the CURVe? command to transfer waveform data.

Table of Contents

Other manuals for Tektronix TDS1002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1002 and is the answer not in the manual?

Tektronix TDS1002 Specifications

General IconGeneral
Bandwidth60 MHz
Channels2
Sample Rate1 GS/s
Record Length2.5k points
Vertical Sensitivity2 mV/div to 5 V/div
Vertical Resolution8 bits
Display Resolution320 x 240 pixels
Display5.7 inch monochrome LCD
Trigger ModesVideo, Edge
ConnectivityUSB 2.0
Weight2.0 kg
Input Impedance1 MΩ
Power Supply100 to 240 VAC, 45 to 440 Hz