Command Descriptions
2−42
TDS200/1000/2000 Series Oscilloscope Programmer Manual
<Event Code><Comma><QString>[<Comma><Event
Code><Comma><QString>...]
<QString>::= <Message>;[<Command>]
<Command> is the command that caused the error and may be
returned when a command error is detected by the oscilloscope. As
much of the command is returned as possible without exceeding the
60 character limit of the <Message> and <Command> strings
combined. The command string is right-justified.
ALLEv?
might return the string :ALLEV 2225,Measurement error, No
waveform to measure; ",420,Query UNTERMINATED; "
*CLS, DESE, *ESE, *ESR?, EVENT?, EVMsg?, EVQty?, *SRE,
*STB?
AUTOSet (No Query Form)
Causes the oscilloscope to adjust its vertical, horizontal, and trigger
controls to display a stable waveform. This command is equivalent
to pushing the front-panel AUTOSET button.
For a detailed description of the Autoset function, refer to the user
manual for your oscilloscope.
Miscellaneous
AUTOSet EXECute
EXECute invokes Autoset.
Examples
Related Commands
Group
Syntax
Arguments

Table of Contents

Other manuals for Tektronix TDS1002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1002 and is the answer not in the manual?

Tektronix TDS1002 Specifications

General IconGeneral
Bandwidth60 MHz
Channels2
Sample Rate1 GS/s
Record Length2.5k points
Vertical Sensitivity2 mV/div to 5 V/div
Vertical Resolution8 bits
Display Resolution320 x 240 pixels
Display5.7 inch monochrome LCD
Trigger ModesVideo, Edge
ConnectivityUSB 2.0
Weight2.0 kg
Input Impedance1 MΩ
Power Supply100 to 240 VAC, 45 to 440 Hz

Related product manuals