Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−61
CURSor:HBArs:UNIts? (Query Only)
Returns the vertical scale units for the selected cursor source
waveform.
Cursor
CURSor:HBArs:UNIts?
VOLts indicates volts from ground as the unit of measure.
DIVs indicates divisions as the unit of measure, with center of screen
as 0 divisions and bottom of screen as –4 divisions.
DECIBELS indicates decibels as the unit of measure, relative to a
1 V
RMS
sine wave (FFT only).
UNKNOWN indicates that Trigger View is active. This also generates
event message 221 (Settings conflict).
CURSor:HBArs:UNIts?
might return CURSOR:HBARS:UNITS VOLTS
CURSor:SELect:SOUrce
Sets or queries the waveform that is the source of the vertical and
horizontal scale factors used in determining cursor values. This
command is equivalent to setting the Source option in the Cursor
menu.
Cursor
CURSor:SELect:SOUrce <wfm>
Group
Syntax
Returns
Examples
Group
Syntax