Math FFT
122
TDS1000/2000-Series Digital Oscilloscope User Manual
The Math FFT function i ncludes three FFT Window options. The re i s
a trade-off between frequency resolution and amplitude accuracy
with each type of window. What you want to measure and your
source signal charact eristics help you to determine which window to
use.
Window Measure Characteristics
Hanning Periodic
waveforms
Better frequency, poorer magnitude
accuracy than Flattop
Flattop Periodic
waveforms
Better magnitude, poorer frequency
accuracy than Hanning
Rectangular Pulses or
transients
Special-purpose window for waveforms that
do not have discontinuities. This is essential-
ly the same as no window
FFT Aliasing
Problems occur when the oscilloscope acquires a time-domain
waveform containing frequency com ponents that are greater than the
Nyquist frequency (refer to Nyquist Frequency on page 117). The
frequency components that a re a bove th e Nyquist frequency are
undersampled, appearing a s lower frequency components that fold
back around the Nyquist frequency. These incorrect components are
called aliases.

Table of Contents

Other manuals for Tektronix TDS1002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1002 and is the answer not in the manual?

Tektronix TDS1002 Specifications

General IconGeneral
Bandwidth60 MHz
Channels2
Sample Rate1 GS/s
Record Length2.5k points
Vertical Sensitivity2 mV/div to 5 V/div
Vertical Resolution8 bits
Display Resolution320 x 240 pixels
Display5.7 inch monochrome LCD
Trigger ModesVideo, Edge
ConnectivityUSB 2.0
Weight2.0 kg
Input Impedance1 MΩ
Power Supply100 to 240 VAC, 45 to 440 Hz

Related product manuals