EasyManua.ls Logo

Tektronix TDS3000 Series - Page 209

Tektronix TDS3000 Series
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix E: Performance Verification
TDS3000 Series User Manual
E–17
9. Set the vertical SCALE to 500 mV/div.
10. Set the horizontal SCALE to 2 ms/div.
11. Adjust the vertical POSITION control to center the time mark
signal on the screen.
12. Adjust the trigger LEVEL as necessary to obtain a triggered
display.
13. Adjust the horizontal POSITION control to move the trigger
location to the center of the screen (50%).
14. Push the DELAY button to turn delay on.
15. Turn the horizontal POSITION control counter-clockwise to set
the delay to exactly 10 ms.
16. Set the horizontal scale to 1 s/div.
17. Check that the rising edge of the marker crosses the center
horizontal graticule line within ±2 divisions of center graticule.
NOTE. One division of displacement from graticule center corre-
sponds to a 100 ppm time base error.
This completes the performance verification procedure.

Table of Contents

Other manuals for Tektronix TDS3000 Series

Related product manuals