17. Press Enter, to exit Calibration mode.
18. Press ↓ button, select Back to Set up menu.
19. Press Enter.
20. Press ↓ button, select Exit and Save Changes.
21. Press Enter.
The Controller will now enter Run mode.
Note, the Auto Lock, Auto Stop, Auto Open and Overcurrent Monitor functions also operate in
the DDA profile.
7.6.4 Calibrating the sash lock function
To define the required Sash Lock position:
1. Press and hold the Enter button for 5 seconds until the Main Menu is displayed.
2. The current Hardware and Software versions are displayed, for example 1. HW2.0 FW1.1
3. Press ↓ button, select 2. Setup and press Enter.
4. Enter the Password using the ↑ button and Enter button, the factory default is 1 0 0 1. In the
Set-Up menu, the controller will display Keypad Tones.
5. Press ↓ button, select Sash Settings and press Enter. The controller will display 1. S1 Enabled.
6. Press ↓ button, select 8. Auto Lock and press Enter.
7. Set to Enabled and press Enter.
8. Use the ↓ button to scroll through the menu and select Lock Alarm. Press Enter.
9. Press ↑/↓ buttons to set the required alarm delay time and press Enter.
10. P Use the ↓ button to scroll through the menu and select Lock Pos Calibration. Press Enter.
11. Instructions will be displayed, for example: Lock the sash then move the sash to the lowest
lock position. Enter to accept. Cancel to abort
12. Open the sash to the Lock position and press Enter.
If the position chosen is too low, lower than the calibrated Bottom, Middle or Top positions,
Bad position - try again will be displayed.
If the chosen position is acceptable, the lock position will be set.
13. Press ↓ button, select Back to Set up menu and press Enter.
14. Press ↓ button, select Exit and Save Changes and press Enter.
The Auto Sash Controller will now go to RUN mode.