EasyManua.ls Logo

Telex C-2000HS - Setup State 3 Guard Tone Level; Setup State 4 Guard Tone Duration; Setup State 5 Hold Tone Level

Telex C-2000HS
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22 Setup C-2000 and C-2000HS Technical Manual
Bosch Security Systems, Inc. Technical Manual F.01U.174.927 Rev 03
Setup State 3 Guard Tone Level
The Guard Tone Level setup state is used to set the level of the guard tone. The guard tone has a 40dB range. The values in
between those shown in Table 5 are valid as well. For example 25 would result in an output level of -6dB for the guard tone.
To set the guard tone level, do the following:
1. Select the setup option using FUP/FDWN keys.
2. Press Enter (#).
Display shows OC (Option Changed).
Setup State 4 Guard Tone Duration
The Guard Tone Duration setup state can vary from 0 to 500ms and is configurable to the nearest 10ms. The setup option is
offset by one (1) and multiplied by 10. Any value between those in Table 6 is valid.
To set the tone duration, do the following:
1. Select the setup option using FUP/FDWN keys.
2. Press Enter (#).
Display shows OC (Option Changed).
Setup State 5 Hold Tone Level
The Hold Tone Level setup state is used to set the level of the hold tone. The Hold tone has a 40dB range. The values in
between those shown in Table 7 are valid as well, for example 25 would result in an output level of -6dB for the hold tone.
To set the hold tone level, do the following:
1. Select the setup option using FUP/FDWN keys.
2. Press Enter (#).
Display shows OC (Option Changed).
TABLE 5. Guard Tone Level
Option Level
1 -30dB
11 -20dB
21 -10dB
25 -6dB
31 0 dB
41 (default) +10dB
TABLE 6. Guard Tone Duration
Option Duration
10 ms
14 (default) 130ms
51 500ms
TABLE 7. Hold Tone Level
Option Level
1 -30dB
11 (default) -20dB
21 -10dB
25 -6dB
31 0 dB
41 +10dB

Table of Contents