6.3. Radio Frequency Protection
6.4. Electro magnetic Compatibility (EMC) Compliance
6.5. Applied Parts
6.5.1. PIXIUM PORTABLE all SERIES
6.6. Biocompatibility
6.6.1. PIXIUM PORTABLE all SERIES
The detector should not be used adjacent to or stacked with other equipment and if adjacent or
stacked use is necessary, the operability of the detector must be tested in the modified
configuration.
The means of communication (mobile phone, walkie talkie, DECT phone...) should be located
outside the patient environment.
The use of radio-frequency transmitters such as portable and mobile Radio Frequency
communication equipments close to the system can affect the operation of the X-Ray Flat-Panel
detector.
The X-Ray Flat-Panel detector should not be used next to or stacked with other equipment.
If this is unavoidable, the detector should be observed to verify correct operation in the
configuration in which it will be used.
Use ESD protection and precautions while unpacking the X-Ray Flat-Panel detector from its
protective bag and installing it in its operational place.
The detector has applied parts: it may be in direct contact with the patient.
The detector is designed to be safe in case of short-term contact with damaged skin (less than 1
hour). Nevertheless, if such case occurs, the detector will be preferably wrapped with sterile
plastic bag.