13. STORAGE AND HANDLING
13.1.PIXIUM PORTABLE EZ-C HD Series Environmental Storage and Transport
Typ min Max
Temperature range - -20 °C +55 °C
Temperature variation - -1 °C/min +1°C/min
Humidity (non-condensing) - 10% RH 75% RH
Pressure (air pressure) - 500 mbar 1500 mbar
Shocks in original package (according to IEC 60068-2-27 Test Ea) - - 25 g
Shocks unpacked (according to IEC 60068-2-27 Test Ea) - - 10 g
Vibrations in original package (sinusoidal vibration according to
IEC 60068-2-6 Test Fc)
- - 10 g
Vibrations unpacked (sinusoidal vibration according to IEC
60068-2-6 Test Fc)
- - 2,5 g
Drop test (All the packed parts of the detector are guaranteed to
resist to a drop test of the following height)
- 70 cm -