12. OPERATING CONDITIONS
12.1.PIXIUM PORTABLE EZ-C HD Series Operating Conditions
Typ Min Max
Temperature +25 °C +0 °C +40 °C
Humidity (for a temperature of 40°C) - 10% RH 75% RH
Pressure - 500 mbar 1500 mbar
Weight (allowed weight uniformly distributed over the complete
surface)
- - 300 kg
Weight (allowed weight uniformly distributed over the surface of a 4
cm diameter disc)
- - 150 kg
Vibrations (The detector shall recover from vibration of at least 2,5
g during 20 minutes, in any directions without any loss in image
quality)
- - 2,5 g