PIXIUM STATIC X-Ray Flat Panel Detectors - End User Manual
Document reference : 63137046 -
January 2024
Page 39
OPEN
This document may not be reproduced, modified, adapted, published, translated, in any way, in whole or in part, or disclosed to a third party without the prior written consent of THALES.
© 2017-2023 THALES All rights reserved
83270259-SER-MIS-EN-002
5) The cleaner should not penetrate into the inside of the PIXIUM EZ HD.
6) Don’t use wipes repeatedly.
7) Don’t spray the cleaner directly on the PIXIUM EZ HD.
8) Don't let the interface get wet.
9) Flat panel detectors should not be soaked in detergent.
10) Don’t wipe too frequently, it may cause damage.
11) After the PIXIUM EZ HD is cleaned, it is dried in time.
Note that the detector has a degree of protection against ingress of water of IP 56 according to IEC 60529.
11.4.PIXIUM all SERIES
The detector housing can be cleaned using a smooth towel, with respect to ESD safety conditions.
Cleaning: wipe the product with a dry, soft, lint-free cloth.
Disinfection: wipe the product with a soft, lint-free cloth slightly moistened in ethylene alcohol or isopropyl alcohol.
Avoid moisture and particles that could contaminate the product connector.
The sterilization, disinfecting, and cleaning conditions are defined by the System manufacturer in the context of the System
use.
11.5.Cleaning Agent
The following table contains clinic agents with the associated active substances.
The detector housing can be cleaned using a smooth towel, with respect to ESD safety conditions.Cleaning: wipe the
product with a dry, soft, lint-free cloth.Disinfection: wipe the product with a soft, lint-free cloth slightly moistened in ethylene
alcohol or isopropyl alcohol.Avoid moisture and particles that could contaminate the product connector.The sterilization,
disinfecting, and cleaning conditions are defined by the System manufacturer in the context of the System use.
11.6.Dark Calibration
When the Pixium EZ HD is going to be cleaned; remove the cables, battery pack from the battery
compartment, and wait until the detector powers off.
Line ID Cleaning Agent
1
Super sani cloth wipes (62% Isopropyl alcohol (IPA) and 0.5% Dual Quaternary Ammonium
Compounds (Quats))
In case of RAD Dual mode, the dark calibration allows acquiring the offset reference for the
preview image correction. This operation has to be performed only if the image acquisition
requires a preview image.
In case of RAD Single mode, the dark calibration allows acquiring the offset reference for the
preview and full X-ray image correction.