11. MAINTENANCE AND CLEANING
In order to ensure the safety and normal use of the equipment, please be sure to check the equipment before use. In the
process of inspection, if any problem is found but cannot be solved, please contact sales representative or product distrib-
utor.
11.1.Daily Inspection
Before using this product, the following checks need to be performed:
11.2.Monthly and yearly inspection
11.3.PIXIUM PORTABLE EZ-C HD
When there are stains on the surface of the PIXIUM EZ HD, such as blood, etc., it needs to be treated. When cleaning a
PIXIUM EZ HD, first the power must be turned off and then it must be cleaned with a non-corrosive cleaner.
1) Turn off the product before cleaning.
2) Using detergent carefully, it avoids damage to the skin and eyes.
3) Cleaning should be done in a ventilated place and gloves should be worn at all times.
4) Don’t use sharp cleaning products.
These cases can occur on detectors whose "power on" function is activated when the cable is
plugged in.The following symptom can be observed when the cable is plugged in, no battery LED
lights up.To recover the detector, turned off it by pressing the power button.Leave the detector on
charge for at least 45 minutes.Once the battery LEDs begin to flash, the detector can be turned on
and used while charging.
Item Operation
CABLE EZ HD
1M
Make sure there is no crack or short circuit on the power supply.
Make sure that the detector cable is not broken or damaged.
Ensure that the power supply cord is reliably connected to the AC input and output socket.
The cable must be firmly connected with the detector without any looseness.
Pixium EZ HD
Make sure that no screws are loose or broken.
Make sure that there is no dust at the connecting pins.
Make sure there is no crack or short circuit at the connecting pins.
Make sure that the Pixium EZ HD can be turned on normally after the power is turned on.
Item Frequency Execution operation
Artifact Monthly / yearly After calibration, check the exposure image.
Calibration
Half a year / when the
exposure condition has
changed
When the number of bad spots or lines of detector
increases, or the core components (such as X-ray
generator, collimator, etc.) are replaced.
BATTERY EZ HD Monthly
Ensure the battery can be charged and discharged
normally.