PIXIUM STATIC X-Ray Flat Panel Detectors - End User Manual
Document reference : 63137046 -
January 2024
Page 41
OPEN
This document may not be reproduced, modified, adapted, published, translated, in any way, in whole or in part, or disclosed to a third party without the prior written consent of THALES.
© 2017-2023 THALES All rights reserved
83270259-SER-MIS-EN-002
11.8.Regular Maintenance
Regular maintenance activity includes the required tests according to IEC 62353.
It must be performed at least every 24 months.
List of activity to be performed on the detector, accessories and parts in this order:Visual check:
Safety related marking, labels and manuals are legible and complete;
Mechanical parts are intact;
There is no damage or contamination;
Documentation is available and reflects the current revision and/or configuration of the detector.
Electrical safety check (according to IEC 62353 recommendations) :
Measure of protective earth resistance;
Measure of device leakage current;
Measure of patient leakage current.
Before launching the calibration, verify that the parameter level.sensitivity in the
DetectorXXXX.ini file is preset with the sensitivity of the detector currently used or PixRad will use
the value got from the detector.