Evolution One Series Spectrophotometers
Spectrophotometer Basics
14 Evolution One Series Spectrophotometer User Guide Thermo Scientific
Related Topics
Purging the Sample Compartment
Purging the Sample Compartment
Purging the sample compartment removes moisture from the sampling area for subambient
temperature-controlled measurements. When purging the instrument, install sample
compartment windows in addition to purge.
Related Topics
Installing Sample Compartment Windows
Selecting a Purge Gas
Installing Purge
Selecting a Purge Gas
Use dry air or dry nitrogen to reduce or eliminate condensation. The purge gas must be free of
moisture, oil and other reactive materials. To remove particulate matter and oil, install a
10-micrometer filter.
Related Topics
Installing Sample Compartment Windows
Installing Purge
Installing Purge
Use 1/4-inch internal diameter tubing to connect a purge gas source to the purge fitting on
the back of the spectrophotometer.
Related Topics
Installing Sample Compartment Windows
Selecting a Purge Gas
WARNING Never use a flammable gas to purge the instrument.